Package: fluke_185_1yr_ver_rs232_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 185: (1 year) CAL VER RS-232 /5520

Date: 2012-05-11 12:38:45

Revision: 1.6

Automated Reference: Fluke 5520A

Model(s) Covered:

185

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5520A 1 year specifications.

VOLTS PEAK HOLD

- ±3.2 Vp; TUR 1.8462

CAPACITANCE

- 4 nF; TUR 2.5568

- 40 nF; TUR 2.4432

- 400 nF; TUR 2.4432

- 4 µF; TUR 2.4432

- 40 mF; TUR 2.4600

AC CURRENT

- 34 µA, 1 kHz; TUR 2.9851

- 4.6 A, 5 kHz; TUR 2.9658

- 10 A, 5 kHz; TUR 3.0556

Source:

Fluke 183/185 Calibration Manual, Part No. 1610436, May 2001,

Rev. 1, 6/02, Manual Supplement 1, Issue Date 3/04

Compatibility:

5500/CAL and MET/CAL 7.3.38 or later

Additional Equipment Required:

DB9 IR Cable Kit, Fluke Part No. 1590638

Capacitance Calibration Cable, Fluke Part No. 1612912