Fluke 18B: (1 year) CAL VER /9100,DMM

Date: 2009-12-16 08:57:15

Revision: 1.1

Automated Reference: Datron 9100

Automated Reference: Digital Multimeter

Model(s) Covered:

18B

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

CAPACITANCE TEST

50 nF Range:

48 nF, TUR 3.3224

500 nF Range:

480 nF, TUR 2.8693

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

ONE 3 V THD LED

Flexible Standard(s) Required:

Digital Multimeter (DMM)

AC Voltage Range: 20 V

Example: Fluke 8808A or equivalent

The average run time during testing was 15 minutes.