Package: fluke_190-202_1yr_ver_rs232_5520_sc600.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 190-202: (1 year) CAL VER RS-232 /5520+SC600, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 190-202: (1 year) CAL VER RS-232 /5520+SC600
Date: 2011-10-18 06:58:17
Revision: 1.1
Automated Reference: Fluke 5520A
Models covered:
190-202
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
INPUT A AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST
900 mV @ 1 MHz; TUR 2.0566
900 mV @ 25 MHz; TUR 3.4803
INPUT B AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST
900 mV @ 1 MHz; TUR 2.0566
900 mV @ 25 MHz; TUR 3.4803
Source:
ScopeMeter 190 Series II
Fluke 190-062, -102, -104, -202, -204, -502 Service Manual
PN 4822 872 05405, July 2011, Rev. 1
Compatibility:
MET/CAL or 5500/CAL 7.3 SP1 (7.3.36) or later
Required Equipment:
See "Sub Fluke 190II VER Required Equipment"
The average run time during testing was 35 minutes in TAR mode.
The average run time during testing was 65 minutes in Meas. Unc. mode.