Package: fluke_190-204_1yr_ver_rs232_5500_sc600.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 190-204: (1 year) CAL VER RS-232 /5500+SC600

Date: 2012-11-07 13:21:17

Revision: 1.5

Automated Reference: Fluke 5500A

Models covered:

190-204

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

INPUT A AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

900 mV @ 25 MHz; TUR 3.4803

INPUT B AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

900 mV @ 25 MHz; TUR 3.4803

INPUT C AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

900 mV @ 25 MHz; TUR 3.4803

INPUT D AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

900 mV @ 25 MHz; TUR 3.4803

Source:

ScopeMeter 190 Series II

Fluke 190-062, -102, -104, -202, -204, -502 Service Manual

PN 4822 872 05405, July 2011, Rev. 2, March 2012

Compatibility:

MET/CAL or 5500/CAL 7.3 SP1 (7.3.36) or later

Required Equipment:

Adapter, dual banana plug to BNC(f), Pomona 1269

Adapter, dual BNC(f) to BNC(m), Fluke PM 9093, 3 ea

Termination, feedthrough, BNC, 50 Ohm, Fluke PM 9585

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-6, 1 ea.

Cable, USB Type A (Host) to USB MINI Type B (Device)

Accessories supplied with UUT

- Power Adapter, BC190

- VPS210 Voltage Probe (200 MHz, 10:1), any color

The average run time during testing was 55 minutes in TAR mode.

The average run time during testing was 105 minutes in Meas. Unc. mode.