Package: fluke_192_1yr_ver_rs232_5500_sc300_5418.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 192: (1 year) CAL VER RS-232 /5500+SC300,5418

Date: 2010-12-14 08:10:51

Revision: 1.7

Automated Reference: Fluke 5500A

Manual Reference: Philips PM 5418

Models covered:

192

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

INPUT A AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

900 mV @ 25 MHz; TUR 3.4803

INPUT B AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

900 mV @ 25 MHz; TUR 3.4803

Source:

Fluke 192/196/199 ScopeMeter Service Manual, Part No. 4822 872 05376,

September 1999, Rev. 2, March 2000

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Required Equipment:

Termination, feedthrough, 75 Ohm, Pomona 4119-75

Cable, 75 Ohm, BNC(m)-(m), Fluke PM 9075

Adapter, dual banana plug to BNC(f), Fluke PM 9081/001

Adapter, dual banana jack to BNC(m), Fluke PM 9082/001

Test leads, stackable, 24", 4 ea.

Termination, feedthrough, 50 Ohm, Fluke PM 9585

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36, 2 ea.

Fluke PM9080 Optically Isolated Interface Cable

Fluke Adapter Kit, Part No. 689486 or equivalent. Includes DB9(f)

to DB9(f) null modem adapter and DB9(m) to DB9(m) gender changer

VPS210-G Voltage Probe (200 MHz, 10:1), Grey or Red