Package: fluke_233_cal_5500.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 233: (1 year) CAL VER /5500

Date: 2009-12-04 13:06:06

Revision: 1.1

Automated Reference: Fluke 5500A

Model(s) Covered:

233

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

CAPACITANCE TEST

100 µF Range:

90 µF, TUR 3.4545

9999 µF Range:

900 µF, TUR 2.0430

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Fluke 80 AK K-type Thermocouple Adapter Accessory

K-type Thermocouple, mini-plug on both ends

The average run time during testing:

Fluke 233: CAL ADJ /5500

Date: 2009-12-04 13:06:12

Revision: 1.1

Automated Reference: Fluke 5500A

Model(s) Covered:

233

Compatibility:

MET/CAL or 5500/CAL SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing: