Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 2686A: (1 year) CAL VER RS-232 /5720,DMM,9346/10k, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 2686A: (1 year) CAL VER RS-232 /5720,DMM,9346/10k
Date: 2011-04-25 09:16:11
Revision: 1.1
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5720A
Manual Reference: Guildline 9346/10k
Model(s) Covered:
2686A
Source:
268XA Data Acquisition System/Data Logging System Service Manual
Document Number: PN 1883791, November 2002
Compatibility:
MET/CAL 7.3 or later
Required Equipment:
Specially configured Universal Input Module
Oscilloscope and its probe
Flexible Standard(s) Required:
Digital Multimeter (DMM)
Voltage Range: 4V
Accuracy: ±0.1%
Example: Fluke 8845A or equivalent
The average run time during testing was 12 minutes (one module).