Package: fluke_287_1yr_ver_rs232_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 287: (1 year) CAL VER RS-232 /5520, Revision: 2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 287: (1 year) CAL VER RS-232 /5520
Date: 2010-08-31 15:06:35
Revision: 2
Automated Reference: Fluke 5520A
Model(s) Covered:
287
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
AC MICROAMPS TEST
500 µA Range: 500.00 µA @ 30 kHz TUR 1.2768
5000 µA Range: 5000.0 µA @ 30 kHz TUR 2.9792
Source:
Fluke 287/289 True-rms Multimeters Calibration Manual
February 2008, Rev. 1, 3/09
Fluke 287/289 True-rms Multimeters Calibration Manual Supplement
Revision/Date 1, 3/09, Issue Date: 6/10, Supplement issue: 3
Fluke USB-IR Cable or USB-RS232 Cable Installation Guide
PN 2395928, February 2005, Rev. 1, 8/06
Compatibility:
MET/CAL or 5500/CAL 7.20 SP3 (7.20Z) or later
Additional Equipment Required:
See "Sub Fluke 287/289 VER Required Equipment"
The average run time during testing was 17 minutes.