Package: fluke_28_ii_2yr_ver_5502.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 28 II: (2 year) CAL VER /5502

Date: 2013-01-29 10:55:58

Revision: 1.1

Automated Reference: Fluke 5502E

Model(s) Covered:

28 II

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5502E (90 day) specifications.

FREQUENCY TESTS (199.99 kHz Range)

199.50 kHz, 150 mV ; TUR 3.9930

RESISTANCE (6 MOhm Range)

3.300 MOhm ; TUR 3.7438

Source:

Fluke 27 II / 28 II Digital Multimeters Calibration Information

April 2010*

Fluke 27 II/28 II Calibration Information Supplement,

Supplement Issue 2, Issue Date July 2012

* The tolerances used by this procedure were taken from the published

specifications table.

Compatibility:

MET/CAL 7.3.38 or later

Additional Equipment Required:

Fluke 80AK or 80AK-A K-Type Thermocouple Adapter

Thermocouple extension cable, Type K, SMP (m)-(m)

The average run time during testing was 9 minutes (keyboard entry).