Fluke 302+: (1 year) CAL VER /5522A,5500COIL, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 302+: (1 year) CAL VER /5522A,5500COIL
Date: 2011-12-15 13:49:58
Revision: 1.2
Automated Reference: Fluke 5500A/COIL
Automated Reference: Fluke 5522A
Model(s) Covered:
302+
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
AC CURRENT:
120.0 A, 50 Hz TUR 2.8850
270.0 A, 50 Hz TUR 2.2222
390.0 A, 50 Hz TUR 2.4384
390.0 A, 150 Hz TUR 2.1352
390.0 A, 400 Hz TUR 2.1352
Source:
Fluke 302+/303 CLAMP METER
Calibration Information
Fluke 302+/303 CLAMP METER
Calibration User Manual
PN 4045130, September 2011
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.3 or later
Additional Equipment Required:
None
The average run time during testing was 6 minutes.