Package: fluke_302_1yr_ver_5502.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 302+: (1 year) CAL VER /5502,5500COIL

Date: 2013-03-13 11:30:18

Revision: 2619

Automated Reference: Fluke 5500A/COIL

Automated Reference: Fluke 5502E

Model(s) Covered:

302+

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1, based on 5502E (90 day) specifications.

AC CURRENT:

120.0 A, 50 Hz TUR 2.8850

270.0 A, 50 Hz TUR 2.2222

300.0 A, 400 Hz TUR 2.0510

390.0 A, 50 Hz TUR 2.4384

390.0 A, 150 Hz TUR 2.1351

Source:

Fluke 302+/303/305 CLAMP METER

Calibration Manual

February 2013

Source Author:

Fluke Calibration

Compatibility:

MET/CAL 8.1.4 or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes.