Package: fluke_312_316_1yr_ver_5520_5500coil.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 312: (1 year) CAL VER /5520,5500COIL, Revision: 1.1
Fluke 316: (1 year) CAL VER /5520,5500COIL, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 312: (1 year) CAL VER /5520,5500COIL
Date: 2009-04-28 07:54:07
Revision: 1.1
Automated Reference: Fluke 5500A-COIL
Automated Reference: Fluke 5520A
Fluke 316: (1 year) CAL VER /5520,5500COIL
Date: 2009-04-28 07:54:10
Revision: 1.1
Automated Reference: Fluke 5500A-COIL
Automated Reference: Fluke 5520A
Model(s) Covered:
316
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
AMPS AC (40 A):
35A @ 50Hz, TUR 1.6078
35A @ 440Hz, TUR 1.5110
AMPS DC (40 A):
-35A, TUR 3.1101
AMPS AC (400 A):
350A @ 50Hz, TUR 2.4864
350A @ 220Hz, TUR 1.7381
AMPS AC (1000 A):
1000A @ 50Hz, TUR 3.6516
Source:
Fluke 312, 316, 318 Clamp Meter
Calibration Manual, PN 2060160 December 2002
Source Author:
Fluke
Compatibility:
5500/CAL or MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 20 minutes.
Model(s) Covered:
312
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
AMPS AC:
35A @ 300Hz, TUR 1.1698
350A @ 50Hz, TUR 4.1440
350A @ 440Hz, TUR 1.3456
900A @ 50Hz, TUR 3.6516
Source:
Fluke 312, 316, 318 Clamp Meter
Calibration Manual, PN 2060160 December 2002
Source Author:
Fluke
Compatibility:
5500/CAL or MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 20 minutes.