Fluke 317: (1 year) CAL VER /9100+Opt200

Fluke 319: (1 year) CAL VER /9100+Opt200

Date: 2009-10-09 13:53:07

Revision: 1.1

Automated Reference: Datron 9100

Model(s) Covered:

317, 319

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

40A AC:

40.00A, 60 Hz; TUR 3.7234

600A AC:

600.0A, 60 Hz; TUR 3.2646

Source:

Fluke 317/319 Clamp Meter User Manual. PN 2672517 Sep 2007

Fluke Verification & Adjustment Information of Houyi (F317 F319)

Rev 0.4, Sep 28, 2007

Source Author:

Fluke

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes or less.

Fluke 317: CAL ADJ /9100+Opt200

Fluke 319: CAL ADJ /9100+Opt200

Date: 2009-10-09 13:53:12

Revision: 1.2

Automated Reference: Datron 9100

Model(s) Covered:

317, 319

Source:

Fluke Verification & Adjustment Information of Houyi (F317 F319)

Rev 0.4, Sep 28, 2007

Source Author:

Fluke

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

The average run time during testing was 6 minutes or less.