Package: fluke_317_319_cal_5520_5500coil.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 317: (1 year) CAL VER /5520,5500A/COIL

Fluke 319: (1 year) CAL VER /5520,5500A/COIL

Date: 2009-10-09 13:51:56

Revision: 1.1

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

317, 319

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

40A AC:

40.0 A, 60 Hz; TUR 1.4682

40A DC:

40.0 A; TUR 2.0579

-40.0 A; TUR 2.0579

600A AC:

30.0A, 60 Hz; TUR 2.2645

600.0A, 60 Hz; TUR 2.1894

600A DC:

600.0A; TUR 2.6706

-600.0A; TUR 2.6706

1000A AC (319 only):

1000A, 60 Hz; TUR 3.0167

1000A DC (319 only):

1000A; TUR 3.5734

-1000A; TUR 3.5734

Source:

Fluke 317/319 Clamp Meter User Manual. PN 2672517 Sep 2007

Fluke Verification & Adjustment Information of Houyi (F317 F319)

Rev 0.4, Sep 28, 2007

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20x or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes or less.

Fluke 317: CAL ADJ /5520,5500A/COIL

Fluke 319: CAL ADJ /5520,5500A/COIL

Date: 2009-10-09 13:52:01

Revision: 1.2

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

317, 319

Source:

Fluke Verification & Adjustment Information of Houyi (F317 F319)

Rev 0.4, Sep 28, 2007

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20x or later

The average run time during testing was 6 minutes or less.