Package: fluke_317_319_cal_5520_5500coil.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 317: (1 year) CAL VER /5520,5500A/COIL, Revision: 1.1
Fluke 319: (1 year) CAL VER /5520,5500A/COIL, Revision: 1.1
Fluke 317: CAL ADJ /5520,5500A/COIL, Revision: 1.2
Fluke 319: CAL ADJ /5520,5500A/COIL, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 317: (1 year) CAL VER /5520,5500A/COIL
Fluke 319: (1 year) CAL VER /5520,5500A/COIL
Date: 2009-10-09 13:51:56
Revision: 1.1
Automated Reference: Fluke 5500A-COIL
Automated Reference: Fluke 5520A
Fluke 317: CAL ADJ /5520,5500A/COIL
Fluke 319: CAL ADJ /5520,5500A/COIL
Date: 2009-10-09 13:52:01
Revision: 1.2
Automated Reference: Fluke 5500A-COIL
Automated Reference: Fluke 5520A
Model(s) Covered:
317, 319
Source:
Fluke Verification & Adjustment Information of Houyi (F317 F319)
Rev 0.4, Sep 28, 2007
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.20x or later
The average run time during testing was 6 minutes or less.
Model(s) Covered:
317, 319
This procedure contains the following test(s) whose Test Uncertainty Ratio
are less than 4:1.
40A AC:
40.0 A, 60 Hz; TUR 1.4682
40A DC:
40.0 A; TUR 2.0579
-40.0 A; TUR 2.0579
600A AC:
30.0A, 60 Hz; TUR 2.2645
600.0A, 60 Hz; TUR 2.1894
600A DC:
600.0A; TUR 2.6706
-600.0A; TUR 2.6706
1000A AC (319 only):
1000A, 60 Hz; TUR 3.0167
1000A DC (319 only):
1000A; TUR 3.5734
-1000A; TUR 3.5734
Source:
Fluke 317/319 Clamp Meter User Manual. PN 2672517 Sep 2007
Fluke Verification & Adjustment Information of Houyi (F317 F319)
Rev 0.4, Sep 28, 2007
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.20x or later
Additional Equipment Required:
None
The average run time during testing was 6 minutes or less.