Fluke 322: (1 year) CAL VER /5080,9100 Opt200,ResBox
Date: 2011-11-10 10:44:09
Revision: 1.1
Automated Reference: Datron 9100 Opt200
Automated Reference: Fluke 5080A
Manual Reference: Resistance Decade
Model(s) Covered:
322
This procedure contains the following test(s) whose Test Uncertainty Ratio
are less than 4:1.
30.0 A, 60 Hz; TUR 2.4880
35.0 A, 50 Hz; TUR 2.5336
50.0 A, 50 Hz; TUR 2.6414
350.0 A, 50 Hz; TUR 2.3776
35.0 A, 400 Hz; TUR 2.5833
50.0 A, 400 Hz; TUR 2.6666
350.0 A, 400 Hz; TUR 2.5000
Source:
Fluke 32X ClampMeters, Calibration Information
Part No. 1631636, October 2001 Rev.1, 12/03.
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.3.10 or later
Additional Equipment Required:
None
The average run time during testing was <tbd> minutes or less.