Fluke 324: (1 year) CAL VER /5502/52120,52120COIL3KA

Date: 2013-03-14 14:36:18

Revision: 2639

Automated Reference: Fluke 52120A

Automated Reference: Fluke 5502E

Model(s) Covered:

324

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5502E (90 day) specifications.

CAPACITANCE TESTS

(1000 µF Range) 500 µF; TUR 2.7692

(1000 µF Range) 900 µF; TUR 2.5743

AC CURRENT:

10.00 A, 50 Hz TUR 3.8368

30.00 A, 50 Hz TUR 3.7419

100.0 A, 50 Hz TUR 3.8368

300.0 A, 50 Hz TUR 3.7419

Source:

Fluke 323/324/325 CLAMP METER

Calibration Manual

February 2013

Source Author:

Fluke Calibration

Compatibility:

MET/CAL 8.1.4 or later

Additional Equipment Required:

Fluke 80AK or 80AK-A K-Type Thermocouple Adapter

Thermocouple extension cable, Type K, SMP (m)-(m)

The average run time during testing was 6 minutes.