Package: fluke_325_1yr_ver_5502.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 325: (1 year) CAL VER /5502,5500COIL

Date: 2013-03-13 11:30:21

Revision: 2622

Automated Reference: Fluke 5500A/COIL

Automated Reference: Fluke 5502E

Model(s) Covered:

325

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5502E (90 day) specifications.

CAPACITANCE TESTS

(1000 µF Range) 500 µF; TUR 2.7692

(1000 µF Range) 900 µF; TUR 2.5743

AC CURRENT:

10.0 A, 50 Hz; TUR 2.9070

30.0 A, 50 Hz; TUR 1.5550

100.0 A, 50 Hz; TUR 3.0864

300.0 A, 50 Hz; TUR 2.5194

DC CURRENT:

10.0 A, 40 A Range TUR 3.5714

30.0 A, 40 A Range TUR 2.2414

100.0 A, 400 A Range TUR 3.9062

300.0 A, 400 A Range TUR 3.2500

Source:

Fluke 323/324/325 CLAMP METER

Calibration Manual

February 2013

Source Author:

Fluke Calibration

Compatibility:

MET/CAL 8.1.4 or later

Additional Equipment Required:

Fluke 80AK or 80AK-A K-Type Thermocouple Adapter

Thermocouple extension cable, Type K, SMP (m)-(m)

The average run time during testing was 6 minutes.