Fluke 333: (1 year) CAL VER /5520,5500A/COIL

Date: 2009-09-29 09:31:37

Revision: 1.3

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

333

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

20 A, 50 Hz; TUR 2.4845

400 A, 50 Hz; TUR 2.1971

600 A, 50 Hz; TUR 2.6616

Source:

Fluke 33X Calibration Manual, Part No. 1618765, June 2001.

Fluke 33x Calibration Maunal Supplement. Part No.1618765, June 2001

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20x or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes or less.

Fluke 334: (1 year) CAL VER /5520,5500A/COIL

Fluke 335: (1 year) CAL VER /5520,5500A/COIL

Date: 2009-09-29 09:31:38

Revision: 1.1

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

334, 335

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

20 A, 50 Hz; TUR 2.4845

200 A, 50 Hz; TUR 2.1971

600 A, 50 Hz; TUR 2.8808

Source:

Fluke 33X Calibration Manual, Part No. 1618765, June 2001.

Fluke 33x Calibration Maunal Supplement. Part No.1618765, June 2001

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20x or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes or less.

Fluke 336: (1 year) CAL VER /5520,5500A/COIL

Fluke 337: (1 year) CAL VER /5520,5500A/COIL

Date: 2009-09-29 09:31:41

Revision: 1.2

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

336, 337

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

20 A, 50 Hz; TUR 2.4845

200 A, 50 Hz; TUR 2.1971

20 A; TUR 3.7488

-20 A; TUR 3.7488

600 A; TUR 3.5139

-600 A; TUR 3.5139

Fluke 337 only.

900 A, 50 Hz; TUR 3.0543

900 A; TUR 3.6367

-900 A; TUR 3.6367

Source:

Fluke 33X Calibration Manual, Part No. 1618765, June 2001.

Fluke 33x Calibration Maunal Supplement. Part No.1618765, June 2001

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20x or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes or less.

Fluke 333: CAL ADJ /5520,5500A/COIL

Fluke 334: CAL ADJ /5520,5500A/COIL

Fluke 335: CAL ADJ /5520,5500A/COIL

Date: 2009-09-29 09:31:47

Revision: 1.1

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

333, 334, 335

Source:

Fluke 33X Calibration Manual, Part No. 1618765, June 2001.

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20x or later

Additional Equipment Required:

DC power supplier: 3 V

The average run time during testing was 6 minutes or less.

Fluke 336: CAL ADJ /5520,5500A/COIL

Fluke 337: CAL ADJ /5520,5500A/COIL

Date: 2009-09-29 09:31:52

Revision: 1.1

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

336, 337

Source:

Fluke 33X Calibration Manual, Part No. 1618765, June 2001.

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20x or later

Additional Equipment Required:

DC power supplier: 3 V

The average run time during testing was 6 minutes or less.