Package: fluke_345_1yr_ver_5080a_9100_opt200.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 345: (1 year) CAL VER /5080,9100 Opt200

Date: 2011-04-14 14:35:39

Revision: 1.3

Automated Reference: Datron 9100 Opt200

Automated Reference: Fluke 5080A

Model(s) Covered:

345

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5080A and 9100 Option 200 (1 year) specifications.

Power Check: (50 Hz and 60 Hz)

Stage 2, Watt @ 40 kW/kVAR Range

379.363 V, 100 A, +73° TUR: 2.0057

Compatibility:

MET/CAL or 5500/CAL 7.3.11 or later

Additional Equipment Required:

None

The average run time during testing was 22 minutes.