Fluke 345: (1 year) CAL VER /5520,5500COIL

Date: 2012-04-05 13:13:13

Revision: 1.4

Automated Reference: Fluke 5500A/COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

345

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5520A (1 year) specifications.

DC Amps Check:

40 A Range

39.00A TUR: 1.8946

400 A Range

41.0A TUR: 3.2303

390A TUR: 2.5823

2000 A Range

1000A TUR: 3.5734

AC Amps Check: (50 Hz and 60 Hz)

40 A Range

39.00A, 60Hz TUR: 1.3480

400 A Range

41.0A, 60Hz TUR: 2.3108

390.0A, 60Hz TUR: 2.0247

2000 A Range

410A, 60Hz TUR: 3.4299

1000A, 60Hz TUR: 3.0167

Power Check: (50 Hz and 60 Hz)

Stage 1, Watt @ 4 kW Range

21 V, 100 A, 0 deg TUR: 3.3367

38 V, 100 A, 0 deg TUR: 3.2062

Stage 1, Watt @ 40 kW Range

380 V, 100 A, 0 deg TUR: 3.2051

Stage 1, VA @ 4 kVA Range

21 V, 100 A, 0 deg TUR: 3.3367

38 V, 100 A, 0 deg TUR: 3.2062

Stage 1, VA @ 40 kW Range

380 V, 100 A, 0 deg TUR: 3.2051

Stage 2, Watt @ 40 kW Range

397.363 V, 100 A, -73 deg TUR: 2.9257

Stage 2, VAR @ 40 kVAR Range

397.363 V, 100 A, -73 deg TUR: 3.1984

Compatibility:

MET/CAL or 5500/CAL 7.3 SP1 (7.3.36) or later

Additional Equipment Required:

None

The average run time during testing was 30 minutes.