Package: fluke_345_1yr_ver_5520a_5500a_coil.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 345: (1 year) CAL VER /5520,5500COIL, Revision: 1.4
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 345: (1 year) CAL VER /5520,5500COIL
Date: 2011-05-11 07:09:39
Revision: 1.4
Automated Reference: Fluke 5500A/COIL
Automated Reference: Fluke 5520A
Model(s) Covered:
345
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
DC Amps Check:
40 A Range
39.00A TUR: 1.8946
400 A Range
41.0A TUR: 3.2303
390A TUR: 2.5823
2000 A Range
1000A TUR: 3.5734
AC Amps Check: (50 Hz and 60 Hz)
40 A Range
39.00A, 60Hz TUR: 1.3480
400 A Range
41.0A, 60Hz TUR: 2.3108
390.0A, 60Hz TUR: 2.0247
2000 A Range
410A, 60Hz TUR: 3.4299
1000A, 60Hz TUR: 3.0167
Power Check: (50 Hz and 60 Hz)
Stage 1, Watt @ 4 kW Range
21 V, 100 A, 0 deg TUR: 3.3367
38 V, 100 A, 0 deg TUR: 3.2062
Stage 1, Watt @ 40 kW Range
380 V, 100 A, 0 deg TUR: 3.2051
Stage 1, VA @ 4 kVA Range
21 V, 100 A, 0 deg TUR: 3.3367
38 V, 100 A, 0 deg TUR: 3.2062
Stage 1, VA @ 40 kW Range
380 V, 100 A, 0 deg TUR: 3.2051
Stage 2, Watt @ 40 kW Range
397.363 V, 100 A, -73 deg TUR: 2.9257
Stage 2, VAR @ 40 kVAR Range
397.363 V, 100 A, -73 deg TUR: 3.1984
Compatibility:
MET/CAL or 5500/CAL 7.3 SP1 (7.3.36) or later
Additional Equipment Required:
None
The average run time during testing was 30 minutes.