Package: fluke_345_1yr_ver_9100_opt200_pwr.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 345: (1 year) CAL VER /9100+Opt200+PWR

Date: 2010-01-08 15:10:56

Revision: 1.3

Automated Reference: Datron 9100

Automated Reference: Datron 9100 Opt200

Model(s) Covered:

345

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

AC Amps Test: (50 Hz and 60 Hz)

40 A Range

39.00A TUR: 3.4511

400 A Range

390.0A TUR: 3.0991

Power Test:

Stage 2, Watt @ 40 kW Range

39.2133 V, 100 A, -73 deg TUR: 3.9270

Stage 2, Watt @ 40 kW Range

46.0105 V, 100 A, 73 deg TUR: 3.3337

397.363 V, 100 A, -73 deg TUR: 1.2861

Stage 2, Watt @ 400 kW Range

460.105 V, 100 A, 73 deg TUR: 2.7230

Compatibility:

MET/CAL 7.20C or later

Additional Equipment Required:

None

The average run time during testing was 30 minutes.