Package: fluke_345_1yr_ver_9100_opt200_pwr.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 345: (1 year) CAL VER /9100+Opt200+PWR, Revision: 1.3
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 345: (1 year) CAL VER /9100+Opt200+PWR
Date: 2010-01-08 15:10:56
Revision: 1.3
Automated Reference: Datron 9100
Automated Reference: Datron 9100 Opt200
Model(s) Covered:
345
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 9100 (1 year) specifications.
AC Amps Test: (50 Hz and 60 Hz)
40 A Range
39.00A TUR: 3.4511
400 A Range
390.0A TUR: 3.0991
Power Test:
Stage 2, Watt @ 40 kW Range
39.2133 V, 100 A, -73 deg TUR: 3.9270
Stage 2, Watt @ 40 kW Range
46.0105 V, 100 A, 73 deg TUR: 3.3337
397.363 V, 100 A, -73 deg TUR: 1.2861
Stage 2, Watt @ 400 kW Range
460.105 V, 100 A, 73 deg TUR: 2.7230
Compatibility:
MET/CAL 7.20C or later
Additional Equipment Required:
None
The average run time during testing was 30 minutes.