Package: fluke_36_1yr_ver_5520_5500coil.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 36: (1 year) CAL VER /5520,5500A/COIL

Date: 2011-06-29 11:51:10

Revision: 1.8

Automated Reference: Fluke 5500A/COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

36

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

DC CURRENT: 200A Range

- 200 A; TUR 2.9932

AC CURRENT: 200A Range

- 200 A, 50 Hz; TUR 2.2023

- 200 A, 60 Hz; TUR 2.2023

- 200 A, 400 Hz; TUR 1.2874

AC CURRENT: 600A Range

- 600 A, 50 Hz; TUR 3.4731

- 600 A, 60 Hz; TUR 3.4731

Source:

Fluke 36 AC/DC True RMS Clamp Meter

Service Information, PN 200409

September 1996, Rev. 1 2/98

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 10 minutes.