Package: fluke_373_1yr_ver_5520_5500coil.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 373: (1 year) CAL VER /5520,5500COIL

Date: 2012-03-15 12:18:11

Revision: 1.2

Automated Reference: Fluke 5500A/COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

373

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

Capacitance Test

100 µF, TUR 3.8182

1000 µF, TUR 3.8182

AC Current Test

20 A TUR 2.4727

100A TUR 2.2023

500A TUR 2.8000

600A TUR 2.8971

Source:

Fluke 373 Clamp Meter Calibration Manual

December 2010

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

None

The average run time during testing was 10 minutes.