Fluke 374: (1 year) CAL VER /5520,5500COIL, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 374: (1 year) CAL VER /5520,5500COIL
Date: 2010-09-16 12:59:29
Revision: 1.1
Automated Reference: Fluke 5500A/COIL
Automated Reference: Fluke 5520A
Model(s) Covered:
374
This procedure contains the following test(s) whose Test Uncertainty Ratio
are less than 4:1.
Capacitance Test
Capacitance Test
500 µF TUR 2.7692
900 µF TUR 2.5743
ROGO Test
300.0 A @ 440 Hz TUR 2.1660
DC AMP Test
500.0 A TUR 3.4863
AC AMP Test
300.0 A, 440 Hz TUR 1.5752
500.0 A, 50 Hz TUR 2.7879
Source:
Fluke sinan f374/5/6 calibration notes
Fluke 374, 375, 376 Clamp Meters User Manual
PN 3608883 July 2010
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.3 or later
Additional Equipment Required:
Flexible Current Probe
- For example, Fluke i2500-10 or i2500-18
The average run time during testing was 10 minutes.