Fluke 376: (1 year) CAL VER /5520,5500COIL

Date: 2010-09-16 12:59:36

Revision: 1.1

Automated Reference: Fluke 5500A/COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

376

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

Capacitance Test

500 µF TUR 2.7692

900 µF TUR 2.5743

ROGO Test

300.0 A @ 440 Hz TUR 2.1660

DC AMP Test

500.0 A TUR 3.4863

900.0 A TUR 3.6367

AC AMP Test

300.0 A, 440 Hz TUR 1.5752

500.0 A, 50 Hz TUR 2.7879

900.0 A, 50 Hz TUR 3.0543

Source:

Fluke sinan f374/5/6 calibration notes

Fluke 374, 375, 376 Clamp Meters User Manual

PN 3608883 July 2010

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.3 or later

Additional Equipment Required:

Flexible Current Probe

- For example, Fluke i2500-10 or i2500-18

The average run time during testing was 10 minutes.