Package: fluke_381_1yr_ver_5520_5500coil.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 381: (1 year) CAL VER /5520,5500COIL, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 381: (1 year) CAL VER /5520,5500COIL
Date: 2010-09-17 07:22:24
Revision: 1.1
Automated Reference: Fluke 5500A/COIL
Automated Reference: Fluke 5520A
Model(s) Covered:
381
This procedure contains the following test(s) whose Test Uncertainty Ratio
are less than 4:1.
DC AMP Test
100.0 A TUR 3.9038
600.0 A TUR 3.5139
-600.0 A TUR 3.5139
1000.0 A TUR 3.6567
-1000.0 A TUR 3.6567
AC AMP Test
100.0 A, 50 Hz TUR 3.0473
500.0 A, 50 Hz TUR 2.8808
975.0 A, 50 Hz TUR 3.0833
ROGO Test
995 A 50 Hz TUR 3.8480
Source:
Fluke 381 Remote Display True-rms Clamp Meter Users Manual
PN 3538357 April 2010
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.3 or later
Additional Equipment Required:
Flexible Current Probe
- For example, Fluke i2500-10 or i2500-18
The average run time during testing was 10 minutes.