Package: fluke_381_1yr_ver_5522_52120_coil3kA_coil6kA.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 381: (1 year) CAL VER /5522,52120,COIL3KA,COIL6KA

Date: 2013-07-05 09:39:27

Revision: 3044

Automated Reference: Fluke 52120A

Automated Reference: Fluke 52120A/COIL3KA

Automated Reference: Fluke 52120A/COIL6KA

Automated Reference: Fluke 5522A

Model(s) Covered:

381

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

DC AMP Test

+100.0 A TUR 3.80

±600.0 A TUR 3.20

+900.0 A TUR 3.34

±975.0 A TUR 3.37

AC AMP Test

100.0 A, 50 Hz TUR 3.84

600.0 A, 50 Hz TUR 3.20

900.0 A, 50 Hz TUR 3.34

975.0 A, 50 Hz TUR 3.36

Source:

Fluke 381 Remote Display True-rms Clamp Meter

Users Manual

PN 3538357 July 2010

Fluke 381 Remote Display True-rms Clamp Meter

Calibration Manual

December 2010 Rev. 1 6/12

Manual Supplement

Manual Title: 381 Calibration

Supplement Issue: 1

Print Date: October 2012

* The tolerances used by this procedure were taken from the published

specifications table.

Additional Equipment Required:

Flexible Current Probe, Fluke i2500-10 or i2500-18

The average run time during testing was 12 minutes.