Fluke 381: (1 year) CAL VER /5522,5500COIL

Date: 2013-06-06 16:06:01

Revision: 3044

Automated Reference: Fluke 5500A/COIL

Automated Reference: Fluke 5522A

Model(s) Covered:

381

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

DC Current

+100.0 A TUR 3.9062

+900.0 A TUR 3.7000

±975.0 A TUR 3.7209

AC Current

100.0 A, 50 Hz TUR 3.0864

600.0 A, 50 Hz TUR 2.9343

900.0 A, 50 Hz TUR 3.1145

975.0 A, 50 Hz TUR 3.1446

Source:

Fluke 381 Remote Display True-rms Clamp Meter

Users Manual

PN 3538357 July 2010

Fluke 381 Remote Display True-rms Clamp Meter

Calibration Manual

December 2010 Rev. 1 6/12

Manual Supplement

Manual Title: 381 Calibration

Supplement Issue: 1

Print Date: October 2012

* The tolerances used by this procedure were taken from the published

specifications table.

Additional Equipment Required:

Flexible Current Probe, Fluke i2500-10 or i2500-18

2 mm to 4 mm Slim reach probe for iFlex, Fluke TP2 PN 650892

Test Lead* for iFlex, Fluke 6358 PN 1903307

*Has 90 ° elbow at one end, to clear probe collar.

The average run time during testing was 14 minutes.