Fluke 434: (1 year) CAL VER RS-232 /5500

Date: 2013-05-14 11:36:28

Revision: 2624

Automated Reference: Fluke 5500A

Models covered:

434

Source:

Fluke 433, 434, 435 Three Phase Power Quality Analyzer

Service Manual, Part No. 4822 872 05396, November 2006;

Service Alert SA-1998, dated Aug 8, 2007.

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on the 5500A 1 year specifications.

VOLTAGE INPUTS ACCURACY : Input A/L1, 230 V Range

460 V @ 50 Hz: TUR 3.7097

VOLTAGE INPUTS ACCURACY : Input B/L2, 230 V Range

460 V @ 50 Hz: TUR 3.7097

VOLTAGE INPUTS ACCURACY : Input C/L3, 230 V Range

460 V @ 50 Hz: TUR 3.7097

VOLTAGE INPUTS ACCURACY : Input N, 230 V Range

460 V @ 50 Hz: TUR 3.7097

Compatibility:

MET/CAL or 5500/CAL 7.3.38 or later

Required Equipment:

See "Sub Fluke 430-Series Cal Required Equipment"

The average run time during testing was 15 minutes.