Package: fluke_434_1yr_ver_rs232_6105a.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 434: (1 year) CAL VER RS-232 /6105A

Date: 2010-11-08 08:05:07

Revision: 1.1

Automated Reference: Fluke 6105A

Models covered:

434

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on the 6105A 1 year specifications.

CURRENT INPUTS ACCURACY : Input A/L1, B/L2, C/L3, N

1000 A Range, 0 A: TUR 2.5000

Source:

Fluke 433, 434, 435 Three Phase Power Quality Analyzer

Service Manual, Part No. 4822 872 05396, November 2006;

Service Alert SA-1998, dated Aug 8, 2007.

Compatibility:

MET/CAL 7.3 or later

Required Equipment:

Adapter, dual banana jack to BNC(m), Fluke PM 9082/001

Adapter, dual BNC(f) to BNC(m), Fluke PM 9093/001, 3 ea.

Test leads, stackable, 24", 4 ea.

Termination, feedthrough, 50 Ohm, Fluke PM 9585, 4 ea.

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-18, 3 ea.

Battery Charger/Power adapter, BC430

FLUKE OC4USB Optical Cable for USB

or Fluke PM9080 Optically Isolated RS-232 cable

The average run time during testing was: