Fluke 434: (1 year) CAL VER RS-232 /9100, Revision: 1.8
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 434: (1 year) CAL VER RS-232 /9100
Date: 2010-12-29 10:10:12
Revision: 1.8
Automated Reference: Datron 9100
Models covered:
434
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on the 9100 1 year specifications.
VOLTAGE INPUTS ACCURACY : Input A/L1, 230 V Range
460 V @ 50 Hz: TUR 3.9249
VOLTAGE INPUTS ACCURACY : Input B/L2, 230 V Range
460 V @ 50 Hz: TUR 3.9249
VOLTAGE INPUTS ACCURACY : Input C/L3, 230 V Range
460 V @ 50 Hz: TUR 3.9249
VOLTAGE INPUTS ACCURACY : Input N, 230 V Range
460 V @ 50 Hz: TUR 3.9249
Source:
Fluke 433, 434, 435 Three Phase Power Quality Analyzer
Service Manual, Part No. 4822 872 05396, November 2006;
Service Alert SA-1998, dated Aug 8, 2007.
Compatibility:
MET/CAL 7.3 later
Required Equipment:
See "Sub Fluke 430-Series Cal Required Equipment"
The average run time during testing was 15 minutes.