Fluke 45: (6 month) CAL VER IEEE /5500

Date: 2009-05-27 14:14:01

Revision: 1.1

Automated Reference: Fluke 5500A

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

DC VOLTAGE TESTS:

100 mV Range, 0.000 mV; TUR 2.0000

100 mV Range, 90.000 mV; TUR 2.8571

300 mV Range, 300.00 mV; TUR 3.8095

AC VOLTAGE TESTS:

300 mV Range, 15.00 mV @ 1 kHz; TUR 3.0588

RESISTANCE TESTS:

3 Mohm Range, 3.0000 Mohm; TUR 3.9604

30 Mohm Range, 30.000 Mohm; TUR 2.5532

300 Mohm Range, 300.0 Mohm; TUR 3.9565

DC CURRENT TESTS:

10 A Range, 10.000 A; TUR 3.9494

AC CURRENT TESTS:

10 A Range, 10.000 A @ 1 kHz; TUR 3.1429

Model(s) Covered:

Fluke 45

Source:

Fluke 45, Dual Display Multimeter Service Manual

PN 609203, March 1999 Rev.1, 1/04

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 6 minutes.

Fluke 45: (1 year) CAL VER IEEE /5500

Date: 2009-05-27 14:14:04

Revision: 1.1

Automated Reference: Fluke 5500A

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

DC VOLTAGE TESTS:

100 mV Range, 0.000 mV; TUR 2.0000

100 mV Range, 90.000 mV; TUR 3.3929

AC VOLTAGE TESTS:

300 mV Range, 15.00 mV @ 1 kHz; TUR 3.0588

RESISTANCE TESTS:

3 Mohm Range, 3.0000 Mohm; TUR 3.9604

30 Mohm Range, 30.000 Mohm; TUR 2.5532

300 Mohm Range, 300.0 Mohm; TUR 3.9565

DC CURRENT TESTS:

10 A Range, 10.000 A; TUR 3.9494

AC CURRENT TESTS:

10 A Range, 10.000 A @ 1 kHz; TUR 3.1429

Model(s) Covered:

Fluke 45

Source:

Fluke 45, Dual Display Multimeter Service Manual

PN 609203, March 1999 Rev.1, 1/04

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 15 minutes.

Fluke 45: CAL ADJ IEEE /5500

Date: 2009-05-27 14:14:06

Revision: 1.1

Automated Reference: Fluke 5500A

Model(s) Covered:

Fluke 45

Source:

Fluke 45, Dual Display Multimeter Service Manual

PN 609203, March 1999 Rev.1, 1/04

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 8 minutes.