Package: fluke_45_cal_ieee_5500.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 45: (6 month) CAL VER IEEE /5500, Revision: 1.1
Fluke 45: (1 year) CAL VER IEEE /5500, Revision: 1.1
Fluke 45: CAL ADJ IEEE /5500, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 45: (6 month) CAL VER IEEE /5500
Date: 2009-05-27 14:14:01
Revision: 1.1
Automated Reference: Fluke 5500A
Fluke 45: (1 year) CAL VER IEEE /5500
Date: 2009-05-27 14:14:04
Revision: 1.1
Automated Reference: Fluke 5500A
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5500A (1 year) specifications.
DC VOLTAGE TESTS:
100 mV Range, 0.000 mV; TUR 2.0000
100 mV Range, 90.000 mV; TUR 3.3929
AC VOLTAGE TESTS:
300 mV Range, 15.00 mV @ 1 kHz; TUR 3.0588
RESISTANCE TESTS:
3 Mohm Range, 3.0000 Mohm; TUR 3.9604
30 Mohm Range, 30.000 Mohm; TUR 2.5532
300 Mohm Range, 300.0 Mohm; TUR 3.9565
DC CURRENT TESTS:
10 A Range, 10.000 A; TUR 3.9494
AC CURRENT TESTS:
10 A Range, 10.000 A @ 1 kHz; TUR 3.1429
Model(s) Covered:
Fluke 45
Source:
Fluke 45, Dual Display Multimeter Service Manual
PN 609203, March 1999 Rev.1, 1/04
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 15 minutes.
Date: 2009-05-27 14:14:06
Revision: 1.1
Automated Reference: Fluke 5500A
Model(s) Covered:
Fluke 45
Source:
Fluke 45, Dual Display Multimeter Service Manual
PN 609203, March 1999 Rev.1, 1/04
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 8 minutes.
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5500A (1 year) specifications.
DC VOLTAGE TESTS:
100 mV Range, 0.000 mV; TUR 2.0000
100 mV Range, 90.000 mV; TUR 2.8571
300 mV Range, 300.00 mV; TUR 3.8095
AC VOLTAGE TESTS:
300 mV Range, 15.00 mV @ 1 kHz; TUR 3.0588
RESISTANCE TESTS:
3 Mohm Range, 3.0000 Mohm; TUR 3.9604
30 Mohm Range, 30.000 Mohm; TUR 2.5532
300 Mohm Range, 300.0 Mohm; TUR 3.9565
DC CURRENT TESTS:
10 A Range, 10.000 A; TUR 3.9494
AC CURRENT TESTS:
10 A Range, 10.000 A @ 1 kHz; TUR 3.1429
Model(s) Covered:
Fluke 45
Source:
Fluke 45, Dual Display Multimeter Service Manual
PN 609203, March 1999 Rev.1, 1/04
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
The average run time during testing was 6 minutes.