Package: fluke_51_52_53_54_ii_cal_525.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 51 II: (1 year) CAL VER /525, Revision: 1.11
Fluke 52 II: (1 year) CAL VER /525, Revision: 1.11
Fluke 53 II: (1 year) CAL VER /525, Revision: 1.11
Fluke 54 II: (1 year) CAL VER /525, Revision: 1.11
Fluke 51 II: CAL ADJ /525, Revision: 1.7
Fluke 52 II: CAL ADJ /525, Revision: 1.7
Fluke 53 II: CAL ADJ /525, Revision: 1.7
Fluke 54 II: CAL ADJ /525, Revision: 1.7
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 51 II: (1 year) CAL VER /525
Fluke 52 II: (1 year) CAL VER /525
Fluke 53 II: (1 year) CAL VER /525
Fluke 54 II: (1 year) CAL VER /525
Date: 2009-07-28 14:02:40
Revision: 1.11
Automated Reference: Fluke 525A
Date: 2009-07-28 14:02:43
Revision: 1.7
Automated Reference: Fluke 525A
Model(s) Covered:
51, 52, 53, & 54 Series II
Source:
51, 52, 53, 54, Series II Thermometer Service Manual
PN 1276123 October 1999
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Type K SMP plug to SMP plug thermocouple cable
Copper SMP plug to SMP plug cable
Fluke 80AK SMP to dual banana adapter
Model(s) Covered:
51, 52, 53, & 54 Series II
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 525A (90 day) specifications.
0 °C; TUR 2.1429
23 °C; TUR 2.2250
-190 °C; TUR 2.7200
1200 °C; TUR 3.0000
Source:
51, 52, 53, 54, Series II Thermometer Service Manual
PN 1276123 October 1999
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later
Sub MET/CAL Version Check
Sub Display Copyright
Sub Select Failed Test Mode
Sub Select Guardbanding & Ver Method (No NTHROW)
Sub Fluke 50 Series II Display & Keypad Tests
Sub Fluke 50 Series II (1 yr) Temperature CAL VER /525
Sub Fluke 53,54 II IR Communication Port Test
Additional Equipment Required:
Type K SMP plug to SMP plug thermocouple cable (1)
IR Interface Adapter, Fluke 53,54 only (2)
IR Cable Adapter Connector, Fluke 53,54 only (2)
1. Included in 5500A/LEADS, Lead Set
2. Only necessary if IR Communication Port is tested.
Included in MET/CAL-SERIAL, Closed Loop Procedure
Serial Connectivity Kit.