Fluke 54 II: (1 year) CAL VER /525

Date: 2012-11-29 15:20:17

Revision: 2.2

Automated Reference: Fluke 525A

Models Covered:

54 Series II

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 525A 1 year specifications.

0 °C; TUR 1.8750

23 °C; TUR 2.2250

-190 °C; TUR 2.7200

1200 °C; TUR 3.3333

Source:

51, 52, 53, 54, Series II Thermometer Service Manual

PN 1276123 October 1999

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Subprocedures:

Sub Display Copyright

Sub Select Failed Test Mode

Sub Select Guardbanding & Ver Method (No NTHROW)

Sub Fluke 54 II Display Test

Sub Fluke 50 Series II Keypad Test

Sub Fluke 50 Series II (1 yr) Temperature CAL VER /525

Sub Fluke 53,54 II IR Communication Port Test

Additional Equipment Required:

Type K extension cable (1)

IR Interface Adapter, Fluke 53,54 only (2)

IR Cable Adapter Connector, Fluke 53,54 only (2)

1. Included in 5500A/LEADS, Lead Set

2. Only necessary if IR Communication Port is tested.

Included in MET/CAL-SERIAL, Closed Loop Procedure

Serial Connectivity Kit.