Fluke 705: (1 year) CAL VER /7526

Date: 2012-10-29 12:28:28

Revision: 1.1

Automated Reference: Fluke 7526A

Model(s) Covered:

705

Source:

Fluke Model 705 Loop Calibrator Calibration Information

PN 667345, December 1998, Rev.1, 12/99

Source Author:

Fluke

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5520A (1 year) specifications.

DC CURRENT SOURCE

4 mA; TUR 2.1429

12 mA; TUR 2.2727

24 mA; TUR 2.0588

DC CURRENT MEASUREMENT

4 mA; TUR 2.5000

12 mA; TUR 3.1250

24 mA; TUR 3.1818

Compatibility:

MET/CAL 8.1.2 or later

Additional Equipment Required: