Package: fluke_705_cal_5520_dmm.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 705: (1 year) CAL VER /5520,DMM, Revision: 1.2
Fluke 705: CAL ADJ /5520, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 705: (1 year) CAL VER /5520,DMM
Date: 2009-09-01 07:45:23
Revision: 1.2
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5520A
Date: 2009-09-01 07:45:28
Revision: 1.1
Automated Reference: Fluke 5520A
Model(s) Covered:
705
Source:
Fluke Model 705 Loop Calibrator Calibration Information
PN 667345, December 1998, Rev.1, 12/99
Source Author:
Fluke
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
None
Model(s) Covered:
705
Source:
Fluke Model 705 Loop Calibrator Calibration Information
PN 667345, December 1998, Rev.1, 12/99
Source Author:
Fluke
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
DC CURRENT MEASUREMENT
24 mA Range: 12.000 mA; TUR 3.4483
24 mA Range: 20.000 mA; TUR 2.6415
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Flexible Standard(s) Required:
Digital Multimeter (DMM)
DC current Range: 100 mA
Accuracy: ±0.0075% or better
Example: Fluke 8508A, HP 3458A
WaveTek/Datron 1281 or equivalent
Additional Equipment Required:
None