Fluke 707Ex: (1 year) CAL VER /5500,DMM

Date: 2012-06-21 12:53:01

Revision: 1.5

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5500A

Model(s) Covered:

Fluke 707Ex

Source:

Fluke 707Ex mA Calibrator Calibration Information

PN 2086601, May 2003 Rev. 2, 5/09

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Required File(s):

user_config_instr.ini

Flexible Standard(s) Required:

Digital Multimeter (DMM):

DC Current: 0 to 26 mA

Accuracy: ±0.004 % ±0.5 µA

Additional Equipment Required:

None

The average run time during testing was 2 minutes.

This procedure contains the following tests whose Test Uncertainty Ratios

are less than 4:1. A guardbanding strategy (chosen by the operator at run

time) is used to compensate for the low TURs.

DC CURRENT MEASUREMENT

- Range 24 mA: 12.000 mA; TUR 2.7586

- Range 24 mA: 24.000 mA; TUR 2.2642