Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 707: (1 year) CAL VER /5520,DMM, Revision: 1.4
Fluke 707Ex: (1 year) CAL VER /5520,DMM, Revision: 1.4
Fluke 707: CAL ADJ /5520,DMM, Revision: 1.1
Fluke 707Ex: CAL ADJ /5520,DMM, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 707: (1 year) CAL VER /5520,DMM
Fluke 707Ex: (1 year) CAL VER /5520,DMM
Date: 2009-09-29 07:48:10
Revision: 1.4
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5520A
Fluke 707Ex: CAL ADJ /5520,DMM
Date: 2009-09-29 07:48:16
Revision: 1.1
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5520A
Model(s) Covered:
707, 707Ex
Source:
Fluke 707 Loop Calibrator
Calibration Information, PN 1631340, November, 2001
Fluke 707Ex mA Calibrator
Calibration Information, PN 2086601 May 2003 Rev.1, 12/03
Source Author:
Fluke
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Required Files:
None
Additional Equipment Required:
None
The average run time during testing was 5 minutes.
Model(s) Covered:
707, 707Ex
This procedure contains the following tests whose Test Uncertainty Ratios
are less than 4:1. A guardbanding strategy (chosen by the operator at run
time) is used to compensate for the low TURs.
Here is list of the TURs less than 4:1:
DC CURRENT MEASUREMENT
- Range 24 mA: 12.000 mA; TUR 2.7586
- Range 24 mA: 24.000 mA; TUR 2.2642
Source:
Fluke 707 Loop Calibrator Calibration Information
PN 1631340, November, 2001
Fluke 707Ex mA Calibrator Calibration Information
PN 2086601, May 2003 Rev. 2, 5/09
Source Author:
Fluke
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Required Files:
user_config_instr.ini
Flexible Standard(s) Required:
Digital Multimeter (DMM):
DC Current: 0 to 26 mA
Accuracy: ±0.004 % ±0.5 µA
Additional Equipment Required:
None
The average run time during testing was 5 minutes.