Fluke 707: (1 year) CAL VER /5520,DMM

Fluke 707Ex: (1 year) CAL VER /5520,DMM

Date: 2009-09-29 07:48:10

Revision: 1.4

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5520A

Model(s) Covered:

707, 707Ex

This procedure contains the following tests whose Test Uncertainty Ratios

are less than 4:1. A guardbanding strategy (chosen by the operator at run

time) is used to compensate for the low TURs.

Here is list of the TURs less than 4:1:

DC CURRENT MEASUREMENT

- Range 24 mA: 12.000 mA; TUR 2.7586

- Range 24 mA: 24.000 mA; TUR 2.2642

Source:

Fluke 707 Loop Calibrator Calibration Information

PN 1631340, November, 2001

Fluke 707Ex mA Calibrator Calibration Information

PN 2086601, May 2003 Rev. 2, 5/09

Source Author:

Fluke

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Required Files:

user_config_instr.ini

Flexible Standard(s) Required:

Digital Multimeter (DMM):

DC Current: 0 to 26 mA

Accuracy: ±0.004 % ±0.5 µA

Additional Equipment Required:

None

The average run time during testing was 5 minutes.

Fluke 707: CAL ADJ /5520,DMM

Fluke 707Ex: CAL ADJ /5520,DMM

Date: 2009-09-29 07:48:16

Revision: 1.1

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5520A

Model(s) Covered:

707, 707Ex

Source:

Fluke 707 Loop Calibrator

Calibration Information, PN 1631340, November, 2001

Fluke 707Ex mA Calibrator

Calibration Information, PN 2086601 May 2003 Rev.1, 12/03

Source Author:

Fluke

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Required Files:

None

Additional Equipment Required:

None

The average run time during testing was 5 minutes.