Package: fluke_712_1yr_ver_5522_8508.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 712: (1 year) CAL VER /5522,8508, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 712: (1 year) CAL VER /5522,8508
Date: 2011-10-21 14:42:56
Revision: 1.2
Automated Reference: Fluke 5522A
Automated Reference: Fluke 8508A
Model(s) Covered:
712
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1, using 5520A/5522A 1 year specifications.
RTD Measure
- 550.00 °C; TUR 3.0000 fw>=1.2,<=1.9
- 300.00 °C; TUR 3.7500 fw>=2.0
Source:
Fluke 71X Series Process Calibrators
Calibration Manual
PN 686540
January 1998 Rev. 10, 5/09
Compatibility:
MET/CAL 7.3 SP1 (7.3.36) or later
Additional Equipment Required:
None
The average run time during testing was 9 minutes (MU mode).