Package: fluke_714_1yr_ver_525.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 714: (1 year) CAL VER /525, Revision: 1.3
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 714: (1 year) CAL VER /525
Date: 2009-03-18 07:24:10
Revision: 1.3
Automated Reference: Fluke 525A
Model(s) Covered:
714
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 525A (90 day) specifications.
Firmware Version < 2.0
THERMOCOUPLE MEASURE:
1200 °C; TUR 3.8889
THERMOCOUPLE SOURCE:
1200 °C; TUR 3.7222
Firmware Version >= 2.0
THERMOCOUPLE MEASURE:
-200 °C; TUR 3.0000
0 °C; TUR 2.8571
800 °C; TUR 2.2222
1200 °C; TUR 2.7778
-10 mV; TUR 3.6923
THERMOCOUPLE SOURCE:
-200 °C; TUR 3.0000
0 °C; TUR 2.8571
800 °C; TUR 2.2222
1200 °C; TUR 2.7778
-10 mV; TUR 3.6923
Source:
Fluke Series 71x Series Process Calibrators
Calibration Manual
PN 686540
January 1998 Rev. 9, 10/08
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Type J SMP plug to SMP plug cable
5500/LEADS set: Thermocouple Extension Type J (1)
Copper SMP plug to SMP plug cable
5500/LEADS set: Thermocouple Extension CU (Copper) (1)
1. Included in 5500/LEADS set.
The average run time during testing was 9 minutes.