Package: fluke_714_1yr_ver_525.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 714: (1 year) CAL VER /525

Date: 2009-03-18 07:24:10

Revision: 1.3

Automated Reference: Fluke 525A

Model(s) Covered:

714

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 525A (90 day) specifications.

Firmware Version < 2.0

THERMOCOUPLE MEASURE:

1200 °C; TUR 3.8889

THERMOCOUPLE SOURCE:

1200 °C; TUR 3.7222

Firmware Version >= 2.0

THERMOCOUPLE MEASURE:

-200 °C; TUR 3.0000

0 °C; TUR 2.8571

800 °C; TUR 2.2222

1200 °C; TUR 2.7778

-10 mV; TUR 3.6923

THERMOCOUPLE SOURCE:

-200 °C; TUR 3.0000

0 °C; TUR 2.8571

800 °C; TUR 2.2222

1200 °C; TUR 2.7778

-10 mV; TUR 3.6923

Source:

Fluke Series 71x Series Process Calibrators

Calibration Manual

PN 686540

January 1998 Rev. 9, 10/08

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Type J SMP plug to SMP plug cable

5500/LEADS set: Thermocouple Extension Type J (1)

Copper SMP plug to SMP plug cable

5500/LEADS set: Thermocouple Extension CU (Copper) (1)

1. Included in 5500/LEADS set.

The average run time during testing was 9 minutes.