Fluke 714: (1 year) CAL VER /5522

Date: 2012-10-29 10:43:43

Revision: 1.2

Automated Reference: Fluke 5522A

Model(s) Covered:

714

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5522A (1 year) specifications.

Firmware Version < 2.0

THERMOCOUPLE MEASURE:

-200 °C; TUR 3.3333

1200 °C; TUR 3.0435

THERMOCOUPLE SOURCE:

-200 °C; TUR 3.4815

1200 °C; TUR 2.9130

Firmware Version >= 2.0

THERMOCOUPLE MEASURE:

-200 °C; TUR 2.2222

0 °C; TUR 2.8571

800 °C; TUR 1.7391

1200 °C; TUR 2.1739

-10 mV; TUR 3.4286

75 mV; TUR 3.1111

THERMOCOUPLE SOURCE:

-200 °C; TUR 2.2222

0 °C; TUR 2.8571

800 °C; TUR 1.7391

1200 °C; TUR 2.1739

-10 mV; TUR 3.4286

75 mV; TUR 3.1111

Source:

Fluke 71x Series Process Calibrators

Calibration Manual

PN 686540

January 1998 Rev. 10, 5/09

Compatibility:

MET/CAL 8.1.2 or later

Additional Equipment Required:

Type J extension cable

5500/LEADS set: Thermocouple Extension Type J (1)

Copper extension cable

5500/LEADS set: Thermocouple Extension CU (Copper) (1)

The average run time during testing was 4 minutes.