Fluke 715: (1 year) CAL VER /5502,DMM

Date: 2012-10-31 12:54:20

Revision: 1.1

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5502E

Model(s) Covered:

715

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5502E (1 year) specifications.

Firmware >= 2.0

DC VOLTAGE MEASURE:

20 V; TUR 3.8095

150 mV; TUR 3.3333

All Firmware

DC CURRENT MEASURE:

24 mA; TUR 2.5660

12 mA; TUR 3.0345

Source:

Fluke 71x Series Process Calibrators

Calibration Manual

PN 686540

January 1998 Rev. 10, 5/09

Compatibility:

MET/CAL 8.1.2 or later

Additional Equipment Required:

None

Flexible Standard(s) Required:

Digital Multimeter (DMM)

DC Voltage Range: 0 to 10 V

Accuracy: 0.005 % + 0.5 mV or better

Example: Fluke 8508A or equivalent

The average run time during testing was 9 minutes.