Fluke 724: (1 year) CAL VER RS-232 /5520,8508, Revision: 1.7
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 724: (1 year) CAL VER RS-232 /5520,8508
Date: 2010-04-01 08:09:33
Revision: 1.7
Automated Reference: Fluke 5520A
Automated Reference: Fluke 8508A
Model(s) Covered: 724
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1. In order to improve this, we use 8508A to characterize
the 5520A in advance for some specified test points whose TUR is lower than
3.5:1. Here is list of TUR before (TUR1) and after characterization (TUR2).
mA Measurement Lower and Upper
- Range 24 mA: 12 mA ; TUR1 3.0345; TUR2 18.333
- Range 24 mA: 24 mA ; TUR1 2.5660; TUR2 3.9799
Source: Fluke 724/725 Calibrators
Calibration Manual
PN 667581, September 1999 Rev.2, 2/06
Compatibility: MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
FLUKE-700SC Serial Interface Cable Assembly (1)
DB9 null modem adapter (1)(3)
DB9 gender changer (1)(3)
2 ea. dual banana to separated single banana shielded
cable. POMONA 1368 A-30 or equivalent
Type K SMP plug to SMP plug thermocouple cable (2)
1. Included in MET/CAL-SERIAL, Closed Loop Procedure
Serial Connectivity Kit.
2. Included in 5500A/LEADS, Lead Set
3. Only needed if UUT is connected to 5520 UUT Serial
Port.