Fluke 725: (1 year) CAL VER RS-232 /5500,8508

Date: 2010-04-01 07:43:47

Revision: 1.5

Automated Reference: Fluke 5500A

Automated Reference: Fluke 8508A

Model(s) Covered: 725

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1. In order to improve this, we use 8508A to characterize

the 5500A in advance for some specified test points whose TUR is lower than

3.5:1. Here is list of TUR before (TUR1) and after characterization (TUR2).

mA Measurement Lower and Upper

- Range 24 mA: 12 mA ; TUR1 3.0345; TUR2 18.333

- Range 24 mA: 24 mA ; TUR1 2.5660; TUR2 3.9788

MILLIAMPS SOURCE: Lower Display

- Range 24 mA; 24 mA ; TUR 3.8549

Source: Fluke 724/725 Calibrators

Calibration Manual

PN 667581, September 1999 Rev.2, 2/06

Compatibility: MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

FLUKE-700SC Serial Interface Cable Assembly (1)

DB9 null modem adapter (1)(3)

DB9 gender changer (1)(3)

2 ea. dual banana to separated single banana shielded

cable. POMONA 1368 A-30 or equivalent

Type K SMP plug to SMP plug thermocouple cable (2)

1. Included in MET/CAL-SERIAL, Closed Loop Procedure

Serial Connectivity Kit.

2. Included in 5500A/LEADS, Lead Set

3. Only needed if UUT is connected to 5500 UUT Serial

Port.