Package: fluke_725ex_1yr_ver_rs232_5520_8508.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 725Ex: (1 year) CAL VER RS-232 /5520,8508, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 725Ex: (1 year) CAL VER RS-232 /5520,8508
Date: 2010-04-01 08:09:42
Revision: 1.1
Automated Reference: Fluke 5520A
Automated Reference: Fluke 8508A
Model(s) Covered: 725Ex
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1. In order to improve this, we use 8508A to characterize
the 5520A in advance for some specified test points whose TUR is lower than
3.5:1. Here is list of TUR before (TUR1) and after characterization (TUR2).
mA Measurement Lower and Upper
- Range 24 mA: 12 mA; TUR1 3.0345; TUR2 18.333
- Range 24 mA: 24 mA; TUR1 2.5660; TUR2 3.9799
Four-Wire RTD: Lower Display
- 550.0 degC; TUR 3.0000
TC Measure: Lower Display
- -180.0 degC: TUR 3.7037
TC Source: Lower Display
- -180.0 degC: TUR 3.7037
MILLIAMPS SOURCE: Lower Display
- Range 24 mA: 20 mA; TUR 2.3810
- Range 24 mA: 24 mA; TUR 3.8549
Source: Fluke 725Ex Multifunction Process Calibrator
Calibration Manual
PN 2406553, January 2005 Rev. 1,5/09
Compatibility: MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
- FLUKE-700SC Serial Interface Cable Assembly
- Fluke 724/725 Calibration Cable, Part No. 1556747
- DB9(f) to DB9(m) null modem and DB9(m) to DB9(m)
gender changer. (Only needed if UUT is connected to
55xx UUT Serial Port.
- Type J SMP plug to SMP plug thermocouple cable
Included in 5500A/LEADS, Lead Set
- MET/CAL-SERIAL, Closed Loop Procedure Serial
Connectivity Kit.