Fluke 726: (1 year) CAL VER RS-232 /5520,8508, Revision: 1.4
Fluke 726: CAL ADJ RS-232 /5520,8508, Revision: 1.4
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 726: (1 year) CAL VER RS-232 /5520,8508
Date: 2009-04-29 09:22:04
Revision: 1.4
Automated Reference: Fluke 5520A
Automated Reference: Fluke 8508A
Fluke 726: CAL ADJ RS-232 /5520,8508
Date: 2009-04-29 09:22:07
Revision: 1.4
Automated Reference: Fluke 5520A
Automated Reference: Fluke 8508A
Model(s) Covered: 726
Source: Fluke 724/726/726 Calibrators
Calibration Manual
PN 667581, September 1999 Rev. 2, 2/06
Fluke 726 Multifunction Process Calibrator
Users Manual
September 2005
Compatibility: MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
FLUKE-700SC Serial Interface Cable Assembly (1), DB9(f) to DB9(f)
null modem and DB9(m) to DB9(m) gender changer (only required if
UUT is connected to 5520 UUT Serial Port) (1)
2 ea., dual banana to separated single banana shielded cable,
Pomona Model 1368 A-30 or equivalent
Type K SMP plug to SMP plug thermocouple cable (2)
1. Included in MET/CAL-SERIAL, Closed Loop Procedure
Serial Connectivity Kit.
2. Included in 5500A/LEADS, Lead Set
Model(s) Covered: 726
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 8508A (1 year) specifications.
MILLIAMPS SOURCE: Lower Display, 24 mA Range
24 mA DC; TUR 2.6123
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
THERMOCOUPLE MEASUREMENT: Upper & Lower Display
0 °C; TUR 2.8571
THERMOCOUPLE SOURCE: Upper & Lower Display
0 °C; TUR 2.8571
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1. In order to improve this, we use 8508A to characterize
the 5520A in advance for some specified test points whose TUR is lower than
3.5:1. Here is list of TUR before (TUR1) and after characterization (TUR2).
MILLIAMPS MEASUREMENT: Lower & Upper Display
4 mA DC; TUR1 3.6923 ; TUR2 11.1627
12 mA DC; TUR1 2.2069 ; TUR2 11.5523
24 mA DC; TUR1 0.9057 ; TUR2 1.4159
Source: Fluke 724/726/726 Calibrators
Calibration Manual
PN 667581, September 1999 Rev. 2, 2/06
Fluke 726 Multifunction Process Calibrator
Users Manual
September 2005
Compatibility: MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
FLUKE-700SC Serial Interface Cable Assembly (1)
DB9 null modem adapter (1)(3)
DB9 gender changer (1)(3)
2 ea. dual banana to separated single banana shielded
cable. POMONA 1368 A-30 or equivalent
Type K SMP plug to SMP plug thermocouple cable (2)
1. Included in MET/CAL-SERIAL, Closed Loop Procedure
Serial Connectivity Kit.
2. Included in 5500A/LEADS, Lead Set
3. Only needed if UUT is connected to 5520 UUT Serial
Port.