Fluke 741B: (1 year) CAL VER RS-232 /5500,3458,LFCTR

Date: 2011-08-29 07:35:52

Revision: 1.3

Automated Reference: Fluke 5500A

Automated Reference: HP 3458A

Automated Reference: Low Frequency Counter

Automated Reference: Low-Frequency Counter

Model(s) Covered:

741B

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (90 day) specifications.

RTD MEASURE (4-WIRE)

780 °C; TUR 3.4783

TC SOURCE

-200 °C; TUR 1.8182

0 °C; TUR 3.1250

800 °C; TUR 1.9231

1370 °C; TUR 1.2500

TC MEASUREMENT

-180 °C; TUR 2.7273

0 °C; TUR 3.1250

800 °C; TUR 2.6923

1300 °C; TUR 2.2500

DC CURRENT MEASUREMENT

20 mA; TUR 2.8889

30 mA; TUR 2.3077

-30 mA; TUR 2.3077

100 mA; TUR 2.1200

-100 mA; TUR 2.1200

Source:

Fluke 74X Series Calibration Manual, Part No. 602505, dated February

1997, Rev. 2 4/99.

Fluke 74X Series Calibration Manual Supplement, Part No. 602505

1997, Rev. 2 4/99.

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Rosemount 644H and programming cable (required for 744 only)

Dual Banana to Dual Banana cable (2)

Single Banana to Single Banana cable (2)

Quality Dual Banana Short

Type K SMP plug to SMP plug thermocouple cable (1)

Fluke cable assembly, PN 949466

DB9 (m) to DB9 (f) straight through cable (exclude 741, 741B)

DB9 (f) connector (exclude 741, 741B)

Flexible Standard(s) Required:

LF Counter (LFCTR):

Accuracy: 1 Hz to 50 kHz, 25 ppm or better

1 Hz to 50 kHz, duty cycle accuracy 1%

Example: Fluke PM 6681 or equivalent

The average run time during testing was: 53 min. (with PM 6681)