Fluke 743: (1 year) CAL VER RS-232 /5500,DMM,LFCTR, Revision: 2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 743: (1 year) CAL VER RS-232 /5500,DMM,LFCTR
Date: 2011-02-16 07:38:43
Revision: 2
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5500A
Automated Reference: Fluke 5520A
Automated Reference: Low-Frequency Counter
Model(s) Covered:
743
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5500A (90 day) specifications.
ACV VOLTAGE MEASUREMENT:
1 V, 40 Hz; TUR 3.1429
RTD MEASURE (4-WIRE)
780 °C; TUR 3.4783
TC SOURCE
-200 °C; TUR 1.8182
0 °C; TUR 3.1250
800 °C; TUR 1.9231
1370 °C; TUR 1.2500
TC MEASUREMENT
-180 °C; TUR 2.7273
0 °C; TUR 3.1250
800 °C; TUR 2.6923
1300 °C; TUR 2.2500
DC CURRENT MEASUREMENT
20 mA; TUR 2.8889
30 mA; TUR 2.3077
-30 mA; TUR 2.3077
100 mA; TUR 1.9925
-100 mA; TUR 2.1200
Source:
Fluke 74X Series Calibration Manual, Part No. 602505, dated February
1997, Rev. 2 4/99.
Fluke 74X Series Calibration Manual Supplement, Part No. 602505
1997, Rev. 2 4/99.
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Rosemount 644H and programming cable (required for 744 only)
Dual Banana to Dual Banana cable (2)
Single Banana to Single Banana cable (2)
Quality Dual Banana Short
Type K SMP plug to SMP plug thermocouple cable (1)
Fluke cable assembly, PN 949466
DB9 (m) to DB9 (f) straight through cable (exclude 741, 741B)
DB9 (f) connector (exclude 741, 741B)
Flexible Standard(s) Required:
Digital Multimeter (DMM)
DC Voltage Range: 14 V
Accuracy: ±0.01% or better
DC Current Range: 22 mA
Accuracy: ±0.01% or better
Example: Fluke 8508A, 1281, HP 3458A or equivalent
LF Counter (LFCTR):
Accuracy: 1 Hz to 50 kHz, 25 ppm or better
1 Hz to 50 kHz, duty cycle accuracy 1%
Example: Fluke 6681 or equivalent
The average run time during testing was: 53 min. (with 8508A, 6681)