Package: fluke_74x_honeywell_2020_1yr_ver_rs232_5520_3458_6680.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 741: (1 year) CAL VER RS-232 /5520,8508,6680

Fluke 741B: (1 year) CAL VER RS-232 /5520,8508,6680

Fluke 743: (1 year) CAL VER RS-232 /5520,8508,6680

Fluke 743B: (1 year) CAL VER RS-232 /5520,8508,6680

Fluke 744: (1 year) CAL VER RS-232 /5520,8508,6680

Honeywell 2020: (1 year) CAL VER RS-232 /5520,8508,6680

Date: 2009-03-17 12:25:39

Revision: 1.6

Automated Reference: Fluke 5520A

Automated Reference: Fluke 8508A

Automated Reference: Philips PM 6680

Model(s) Covered:

741, 741B, 743, 743B, 744, Honeywell 2020

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5520A (90 day) specifications.

TC SOURCE

-200 °C; TUR 2.4000

800 °C; TUR 2.6316

1370 °C; TUR 1.6667

TC MEASUREMENT

780 °C; TUR 3.8095

-180 °C; TUR 3.6000

800 °C; TUR 3.6842

1300 °C; TUR 3.0000

200 °C; TUR 2.7673

DC CURRENT MEASUREMENT

20 mA; TUR 3.6111

30 mA; TUR 2.8846

100 mA; TUR 2.6500

Source:

Fluke 74X Series Calibration Manual, Part No. 602505, dated February

1997, Rev. 2 4/99.

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Rosemount 644H and programming cable (required for 744 only)

The average run time during testing was: 53 min. (with 8508A ACAL ALL)

38 min. (w/o 8508A ACAL ALL)