Package: fluke_74x_honeywell_2020_1yr_ver_rs232_5520_3458_6680.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 741: (1 year) CAL VER RS-232 /5520,8508,6680, Revision: 1.6
Fluke 741B: (1 year) CAL VER RS-232 /5520,8508,6680, Revision: 1.6
Fluke 743: (1 year) CAL VER RS-232 /5520,8508,6680, Revision: 1.6
Fluke 743B: (1 year) CAL VER RS-232 /5520,8508,6680, Revision: 1.6
Fluke 744: (1 year) CAL VER RS-232 /5520,8508,6680, Revision: 1.6
Honeywell 2020: (1 year) CAL VER RS-232 /5520,8508,6680, Revision: 1.6
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 741: (1 year) CAL VER RS-232 /5520,8508,6680
Fluke 741B: (1 year) CAL VER RS-232 /5520,8508,6680
Fluke 743: (1 year) CAL VER RS-232 /5520,8508,6680
Fluke 743B: (1 year) CAL VER RS-232 /5520,8508,6680
Fluke 744: (1 year) CAL VER RS-232 /5520,8508,6680
Honeywell 2020: (1 year) CAL VER RS-232 /5520,8508,6680
Date: 2009-03-17 12:25:39
Revision: 1.6
Automated Reference: Fluke 5520A
Automated Reference: Fluke 8508A
Automated Reference: Philips PM 6680
Model(s) Covered:
741, 741B, 743, 743B, 744, Honeywell 2020
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (90 day) specifications.
TC SOURCE
-200 °C; TUR 2.4000
800 °C; TUR 2.6316
1370 °C; TUR 1.6667
TC MEASUREMENT
780 °C; TUR 3.8095
-180 °C; TUR 3.6000
800 °C; TUR 3.6842
1300 °C; TUR 3.0000
200 °C; TUR 2.7673
DC CURRENT MEASUREMENT
20 mA; TUR 3.6111
30 mA; TUR 2.8846
100 mA; TUR 2.6500
Source:
Fluke 74X Series Calibration Manual, Part No. 602505, dated February
1997, Rev. 2 4/99.
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Rosemount 644H and programming cable (required for 744 only)
The average run time during testing was: 53 min. (with 8508A ACAL ALL)